Portable pH/mV Meter - HI8424
SKU: HI8424
The HI8424 General Purpose pH/mV Meter is a highly accurate compact, waterproof, portable meter with a separate temperature probe for temperature compensated pH measurement.This meter has a BNC connection and can be used with many BNC style pH and ORP electrodes.
- Compact waterproof design
- Automatic Temperature Compensation (ATC)
- Automatic one or two-point calibration
Features
Waterproof, heavy duty casing protects against water that would damage the electronics
Automatic calibration to one or two points to standard buffer values (pH 4.01, 7.01, and 10.01)
Automatic shutoff after 20 minutes to conserve battery life.
Low battery indicator is displayed when the battery life is at 5%
High resolution/accuracy mV range, switching resolution automatically from 1 mV to 0.1 mV when ORP readings are between -699.9 to +699.9 mV. The accuracy within this range is ±0.2 mV.
Comes complete with a protective carrying case, PEI plastic body pH electrode, stainless steel temperature probe, calibration buffer sachets, and cleaning solution sachets
SKU | HI8424 |
---|---|
Product Name | Portable pH/mV Meter - HI8424 |
Quote Required | No |
pH Range | -2.00 to 16.00 pH |
pH Resolution | 0.01 pH |
pH Accuracy | ±0.01 pH |
pH Calibration | automatic, at one or two points with three standard buffers (4.01, 7.01,10.01) |
pH Input impedence | 10¹² Ohms |
mV Range | ±699.9 mV; ±1999 mV |
mV Resolution | 0.1 mV; 1 mV |
mV Accuracy | ±0.2 mV; ±1 mV |
Temperature Range | -20.0 to 120.0°C / -4.0 to 248.0°F |
Temperature Resolution | 0.1°C / 0.1°F |
Temperature Accuracy | ±0.4°C; ±0.8°F |
Temperature Compensation | automatic from -20.0 to 120.0ºC (-4.0 to 248.0ºF) or manual without temperature probe connected |
pH Electrode | HI1230B PEI body pH probe with BNC connector and 1 m (3.3’) cable (included) |
Temperature Probe | HI7662 stainless steel temperature probe with 1 m (3.3’) cable (included) |
Automatic Shut-Off | after 20 minutes of non-use (can be disabled) |
Battery Type/Life | 9V / approximately 150 hours of continuous use |
Environment | 0 to 50°C (32 to 122°F); RH max 100% |
Dimensions | 164 x 76 x 45 mm (6.5 x 3.0 x 1.8”) |
Weight | 180 g (6.3oz.) |
Ordering Information | HI8424 is supplied with HI1230B pH electrode, HI7662 temperature probe, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachets (2), battery, protective case and instructions. |